Micromeritics Subsieve AutoSizer (SAS)
The SAS is designed to generate “Fisher number” results identical to that of its predecessor (the Fisher FSSS). This is a crucial point, as the air permeability technique and the FSSS have been used as a benchmark for decades in many industries. Many applications use historical data and quality control standards, and a modern source of comparable, repeatable results is necessary.
Specifications:
Height: 55 cm (21.7 in)
Width: 50 cm (19.7 in)
Depth: 38 cm (15 in)
Weight: 28 kg (61.7 lb)
SIZE RANGE: 0.5 – 75 micron
Voltage: 120 – 240 VAC
Frequency: 50 – 60 Hz
Current: 1A
POROSITY RANGE: 0.2–0.9
COMPRESSION ACCURACY: <0.05mm


What is Air-permeability Particle Sizing?
The SAS utilizes dual pressure transducers to measure pressure drop across a packed bed of powder. By varying the sample height and porosity while controlling the flow rate of air through the sample, the SSA and average particle size can be determined using the Kozeny-Carman equation.
Features / Benefits:
Direct Comparison of SAS and FSSS Results: